XgardIQ (Please call for price)
Overview
XgardIQ is an intelligent and versatile fixed detector and transmitter compatible with Crowcon’s full range of sensor technologies. Available fitted with a variety of sensors for fixed flammable, toxic, oxygen or H2S gas detection.
Providing analogue 4-20mA and RS-485 Modbus signals as standard, XgardIQ is optionally available with Alarm and Fault relays and HART communications. The 316 stainless steel are available with three M20 or 1/2“NPT cable entries.
(SIL-2) Safety integrity level 2 certified fixed detector.
Easy and safe to use
- OLED Display – Bright, clear display indicates gas level and detector status even in dark locations
- Non-Intrusive Calibration – Zero and calibration functions are performed via the display and keypad, without the need for a hot-work permit or any special tools (e.g. no magnetic wand required)
- +Ve Safety Indicator – Confirms the detector is operating safely at a glance
- Event Log Function – Stores all alarm, fault and maintenance events to provide a history of detector use
- Remote Sensor – Improves probability of detection
Simple installation and maintenance
- Universal Transmitter – Common transmitter for Crowcon’s full range of sensor technologies
- Hot-Swap Sensor Modules – Sensors can easily be replaced using one hand without requiring tools or a hot-work permit
- Auto-Sense Function – Automatically detects whether the control system is 4-20mA current sink or source and sets itself appropriately, eliminating the risk of being incorrectly set
- Auto-Configure Function – Detects when a sensor module is plugged-in and uploads the appropriate gas type, range, units and alarm levels
Reliability
- Smart Bump And Speedy Bump Test Functions – Enable sensor health and response to be verified quickly and easily using test gas. Process is automated and provides a reminder when a bump test is due
- Calibration Due Warning – Reminds the user when calibration is due to ensure that sensors remain operational and accurate at all times
- Sensor Range Selection – User can set the full-scale range of the sensor via the display menu according to site or specific location requirements.
Specifications
Specifications
Size H278 x W140 x D89mm (10.9 x 5.5 x 3.5 in) Weight 4.1Kg (9lbs) Enclosure material 316 stainless steel Ingress protection IP66 Connection Three M20 or ½”NPT cable gland entries.Certified, removable plugs are fitted to left-hand and lower righthand entries Power 14-30 VDC. <4W Display Main display – OLED 128 x 64 pixels, yellow text on black background
Indicators – Amber, Red and Green LED’s for detector status; Blue +ve Safety LEDElectrical output 4-20mA current sink or source (Auto-Sense or manual selection); warning and fault signals are configurable. NAMUR NE 43 compliant
RS-485 Modbus RTU
HART 7 over 4-20mA signal and via local I.S. test points (optional)
Relays (optional)
Alarm 1, Alarm 2, Fault
SPCO contacts rated 5A, 230Vac
Relay configuration options
Energised or de-energised
Latching or non-latching
Rising or falling
Configurable On and Off delays on alarm relays.Event logging Records alarm, fault and maintenance events. Events can be viewed on-screen and downloaded to a PC. Operating temperature Transmitter only: -40°C to +75°C (-40°F to +167°F)† Sensor type:
Oxygen: (-20°C to +50°C ambient)
Toxic: (-40°C to +50°C ambient)
Pellister: (-40°C to +75°C ambient)
Infra-red (IR): (-20°C to +55°C ambient)
High temperature toxic: (-40⁰C to +70⁰C)Humidity 0 to 95% RH non-condensing† Repeatability +/- 2% FSD Zero drift +/- 2% FSD per year maximum Response time Sensor dependant: contact Crowcon for specific sensor data. Approvals ATEX and IECEx:
Ex II 2 G Ex db ia IIC T4 Gb (see above for temperature ranges)Performance Designed for compliance with:
EN50104 (oxygen sensors)
EN45544 (toxic gas sensors)*
EN50271 (software)
H2S performance standardFunctional safety IEC61508, EN50402 SIL 2 EMC Compliance EN50270:2015
FCC CFR47 Part 15B
ICES-003† Note: sensor operating temperature and humidity limits vary widely.
* Features outstanding at the time of issue, contact Crowcon for details.
Gases and Ranges
Gas Units Max
RangeStandard
RangeAlarm 1
levelAlarm 2
levelTemp
range °COxygen (O2) %Vol 30 25 19 (falling) 17 (falling) -20 to +50 Ammonia PPM 100/1000 50 25 35 -20 to +40 Carbon Monoxide (CO) PPM 1000 250 30 200 -30 to +50 Hydrogen Fluoride (HF) PPM 10 10 1.8 3 +5 to +40 Hydrogen Sulphide (H2S) PPM 100/200 25 5 10 -30 to +50 High Temperature Hydrogen Sulphide (H2S) PPM 100/200 25 5 10 -40 to +70 Ozone (O3) PPM 1 1 0.1 0.2 -20 to +40 Sulphur Dioxide (SO2) PPM 100 10 1 1 -30 to +50 Hydrogen (H2) %LEL 100 100 20 40 -40 to +75 Methane (CH4) %LEL 100 100 20 40 -40 to +75 Pentane (C5H12) %LEL 100 100 20 40 -40 to +75 Propane (C3H8) %LEL 100 100 20 40 -40 to +75 Ethanol (IR sensor) %LEL 100 100 20 40 -20 to +55 LPG (IR sensor) %LEL 100 100 20 40 -20 to +55 Methane (IR sensor) %LEL 100 100 20 40 -20 to +55 Propane (IR sensor) %LEL 100 100 20 40 +10 to + 55 Delivery & Returns